7.3
General Rules for Maintenance and Service
General Rules
Opening Flow Path Connections
Depot Repair
Charged Aerosol Detectors (VH-D20, VF-D20)
Operating Manual
For successful maintenance and service procedures, follow the rules and
recommendations below.
•
Before starting maintenance or service procedures, shut down the
device when instructed to do so.
•
Use only the replacement parts specifically authorized and qualified
for the device by Thermo Fisher Scientific.
•
Follow all instructions step by step and use the tools recommended
for the procedure.
•
Before opening the flow path to replace capillaries in the system,
turn off the pump flow and wait until the system pressure is down to
zero.
•
Dirty components can contaminate the chromatography system.
Contamination leads to poor performance of the modules and entire
system or can even cause damage to the modules and system.
Therefore:
Always wear appropriate gloves.
¨
Place the components only on a clean, lint-free surface.
¨
Keep your tools clean.
¨
Use only lint-free cloth for cleaning.
¨
•
If you need to return the device for depot repair, follow the
instructions in
Transporting or Shipping the Device
See also
2
Consumables and Replacement Parts (} page 177)
7
•
Maintenance and Service
(} page 145).