Thermo Scientific VH-D20 Operating Manual page 70

Charged aerosol detectors
Hide thumbs Also See for VH-D20:
Table of Contents

Advertisement

5
Installation
Page 70
5. On the right side of the detector, connect the elbow connectors.
a) On the left port (gas inlet), push the connection stem of the gas
inlet elbow connector into the port.
b) On the right port (gas exhaust), push the connection stem of the
gas exhaust elbow connector into the port.
c) Slightly pull on each elbow connector to check that they are
properly locked in the ports. If an elbow connector slips out of a
port, push it a little bit further into the port and repeat the
check.
Figure 17: Pushing in the connection unions
No.
Description
1
Gas inlet port
2
Gas inlet elbow connector with gas inlet tubing
3
Gas exhaust port
4
Gas exhaust elbow connector with gas exhaust tubing
Charged Aerosol Detectors (VH-D20, VF-D20)
Operating Manual