5
Installation
5.5.3

Connecting the Interface Cables

Page 60
The procedure depends on the system configuration:
System configuration
The charged aerosol
detector is the only
detector in the Vanquish
system
The charged aerosol
detector is the second
detector in the Vanquish
system, for example, after
a diode array detector
This picture shows the USB and system interlink connections if the
charged aerosol detector is the second detector in the system, for
example, after a diode array detector.
Follow these steps
Follow the instructions in the Vanquish System
Operating Manual to set up the USB and system
interlink connections in the system.
Set up the USB and system interlink connections as
follows (see picture below this table for the complete
USB and system interlink connections):
1. Follow the instructions in the Vanquish System
Operating Manual to connect the other modules in
the system up to the diode array detector.
2. Connect the "B" connector of a USB cable to the USB
port on the charged aerosol detector, and connect
the "A" connector of the USB cable from the charged
aerosol detector to the USB hub on the diode array
detector.
3. Connect the "A" connector of the USB cable from the
diode array detector to a USB port (2.0 or higher) on
the computer.
4. Connect one end of the system interlink cable to the
free System Interlink port on the diode array
detector, and the other end to a System Interlink port
on the charged aerosol detector.
5. Connect one end of the system interlink cable to the
free System Interlink port on the charged aerosol
detector, and the other end to a System Interlink port
on the column compartment.
Charged Aerosol Detectors (VH-D20, VF-D20)
Operating Manual