Exposure and Measurement Inspection

1
Press a on the touch panel, and select [MANUAL].
• The selected item illuminates green. Each press switches the
item between [AUTO] and [MANUAL].
2
Press UV of [BRIGHTNESS] and check that
the level of the indicator changes.
3
Press a, and select [AUTO].
• The selected item illuminates green. Each press switches the
item between [AUTO] and [MANUAL].
• Selection is not possible when an endoscope is not connected.
4
Press b, and select [AVERAGE] or [PEAK].
• The selected item illuminates green. Each press switches the item between [AVERAGE] and [PEAK].
• Selection is not possible when an endoscope is not connected.

Color Adjustment

1
Adjust the white balance before adjusting the color balance.
4
I "White Balance Button" (P.27)
2
Press c on the touch panel.
• The color balance adjustment screen appears.
3
Press UV of [BLUE] or [RED] and check the
level value. Gently place your hand around the distal
end of the endoscope, and check the changes in the
image of your hand on the monitor.

Suction Inspection

1
Place the distal end of the endoscope in a basin of
sterile water, and press the suction control valve.
(Refer to Figure 4.3.)
• The sterile water should be rapidly aspirated into the suction
system.
NOTE
Be sure to use freshly distilled or sterilized water.
21
Pre-use Inspections
Suction
Figure 4.3