4

Pre-use Inspections

To ensure the equipment is in good working condition for use with patients, check that the endoscope, processor, and
accessories are clean and operate properly before use. Also, provide spare equipment in case use of the equipment must be
stopped as result of a malfunction or other unforeseen problem.
Be sure to perform the following inspection steps before using the processor. If you discover any abnormality, stop using the
processor immediately and contact a PENTAX Medical service facility.
Check the following before you begin the inspection.
• The processor is turned off.
• The processor is installed on a stable and level surface.
• The water bottle assembly is prepared properly and connected correctly.
• The power cord is connected properly.
• The endoscope is connected properly.
• The keyboard is connected properly.
• The USB flash memory or other storage media is connected properly.
Before saving any images, check that
(a) the format for the data to be saved to the USB flash memory is the desired format, and
(b) the USB flash memory or similar is selected as the storage location.
I "Select drive (Still image)" (P.60), "Select drive (Video image)" (P.63)
WARNING
Be sure to perform the following inspections before you use the processor. In the event that any function or device in the
4
video endoscope system does not work properly, do not perform the endoscopic examination. Contact a PENTAX
Medical service facility before using the equipment.

Power-On and Lamp Inspection

1
Turn on the monitor and peripheral devices.
2
Press the R power switch to turn on the processor, and check that the R LED illuminates.
• The touch panel turns on.
• Check that the LIFE indicator (lamp life indicator) on the touch panel is lit green.
• If pressing R does not turn on the processor, check the fuses.
I "Replacing the Fuses" (P.84)
 The lamp life is 500 hours. Check the LIFE indicator (lamp life indicator) on the touch panel before using the
processor. Replace the lamp cartridge if one red circle on the LIFE indicator is lit red and a message prompting you
to replace the lamp is displayed on the monitor. The lamp life may not be as long as 500 hours depending on the
method of use and operating environment.
I "Replacing the Lamp" (P.82)
 If the touch panel does not lit up after more than one and a half minutes from the system startup, restart the system.
3
Press S on the touch panel.
• S illuminates green.
4
Check that the main lamp of the processor illuminates, and the distal end of the endoscope emits light.
WARNING
 Do not look directly at the light emitted from the distal end of the endoscope. Doing so can cause eye injury.
 Do not open the lamp housing cover to avoid electric shock.
NOTE
 If the lamp fails to illuminate, stop the inspection and contact a PENTAX Medical service facility.
 If a PENTAX Medical video endoscope or video module that is not compatible with the processor is connected,
endoscopic images will not be displayed.
 The auxiliary lamp is for the purpose of ensuring that you can see what you are doing while removing the endoscope
in the event of the main lamp fails during a procedure. If this occurs, release the angulation of the endoscope, remove
the endoscope from the patient's body cavity, and stop the examination. Do not perform a procedure using the
auxiliary lamp.