 i-scan setting
Specify the i-scan modes that can be switched using a customize button on the touch panel, endoscope button, foot switch, or
keyboard.
1
Press [>].
• The [i-scan setting] screen appears.
• The mode that is currently specified is indicated by green
characters at the top of the screen.
2
Press the buttons of the modes you want to use.
• The selected mode buttons illuminate blue.
• Pressing a button again cancels the selection.
3
Press [OK].
• If you switch to another screen without pressing [OK], the
settings will not be applied.
 Freeze scan
Set whether or not to use the freeze scan function.
[Off]:
The image is displayed as is when the freeze function is executed.
[Short][Middle][Long]:
The image with the least blur is selected automatically from the images up to a few seconds before
the freeze function was executed.
The amount of time for image selection increases in the following order:[Short],[Middle], and [Long].
5
1
Press [Off],[Short],[Middle] or [Long].
NOTE
When you use the freeze scan function, wait for the endoscopic image on the LCD monitor to show the effect of desired
image processing before capturing the image.
57
Operating Procedure
The mode that is currently specified is
indicated by green characters.
Select the modes you want to use.
Confirm the selection.