Using OE
OE modes can be activated during i-scan or white light observations.
Switching from White light image to OE image:
1
If i-scan functions are activated: Press [i-scan1], [i-scan2], [i-scan3] key on the keyboard, or
customize button on the touch panel, or configure setting on [i-scan] Tab.
• Select OFF of i-scan1, 2 or 3.
I "Function Keys" (P.73), "Customize button settings" (P.44), "[i-scan] Tab" (P.35)
2
Press the [OE] key on the keyboard / customize button on the touch panel or configure setting
on [i-scan] Tab.
• Each press changes the mode of OE.
• Select from [Mode1], [Mode2] and [off].
I "Function Keys" (P.73), "Customize button settings" (P.44), "[i-scan] Tab" (P.35)
Switching from i-scan1, i-scan2, or i-scan3 image to OE image:
1
Press the [OE] key on the keyboard / customize button on the touch panel or configure setting
on [i-scan] Tab with i-scan1, 2 or 3 function are been activating.
• Each press changes the mode of OE.
• Select from [Mode1], [Mode2] and [off].
I "Function Keys" (P.73), "Customize button settings" (P.44), "[i-scan] Tab" (P.35)
NOTE
70K and 80K series endoscopes are not compatible with OE, therefore OE function cannot be used.
90K, K10 and J10 series endoscopes are not compatible with Twin mode viewing in OE mode image observation.
Some of 90K series endoscope require software update to be used with OE. Please contact your local PENTAX
Medical service facility.
Using Enhancement
Configure the edge enhancement setting.
1
Press the [Enhance level] key on the keyboard / customize button on the touch panel.
• Each press changes the intensity of enhancement.
I "Function Keys" (P.73), "Customize button settings" (P.44)