Automatic Brightness Control Inspection
Check the video images from the scope on the monitor while the lamp is lit.
1
Bring the distal end of the scope to about 1 cm from
the palm of your hand and then slowly move it away
until it is about 5 cm from your hand while making
sure the brightness of your hand displayed on the
monitor is uniform. (Refer to Figure 3.1.)
3
Scope Control Button Inspection
1
Press each button on the control body of the
endoscope and make sure a confirmation tone
sounds. (Refer to Figure 3.2.)
• For details on each of the buttons on the control body of the
endoscope, refer to the instruction manual for the scope.
Exposure and Measurement Inspection
1
Press Y on the touch panel and select
[MANUAL].
• The selected item lights green. Each press switches the item
between [AUTO] and [MANUAL].
2
Press UV of [Brightness] and make sure the
indication of the indicator changes.
3
Press Y and select [AUTO].
• The selected item lights green. Each press switches the item
between [AUTO] and [MANUAL].
• Selection is not possible when a scope is not connected.
4
Press Z and select [AVE] or [PEAK].
• The selected item lights green. Each press switches the item between [AVE] and [PEAK].
• Selection is not possible when a scope is not connected.
18
Pre-use Inspections
Figure 3.1
Figure 3.2