Message Displayed on
Monitor
System error. Please restart
the processor(No**-**).
System error. Please contact
your PENTAX service
facility(No**-**).
AUX lamp ON. Please contact
your PENTAX service facility.
Copy failed.
Data transfer failed.
PC:Not ready.
PC:HDD alert.
PC:HDD warning.
PC:Transfer error.
IP CONFLICT
Video rec error.
7
No data.
Changing OE mode failed!
Please retry.
Changing OE mode Off failed!
Check scope
connection.(No.**)
Out of internal memory space.
89
Troubleshooting
Meaning
There was a problem with the system
startup.
There was a problem with the system
startup.
Either the lamp was turned on when an
illumination function error occurred, or
an illumination function error occurred
when turning on the lamp.
Copying the procedure history or
system configuration to Image
Recording Media failed.
Failed to transfer data to peripheral
devices.
Capture is done while unconnected to
Endoimage2.
There is no sufficient space at PC in
which installed Endoimage2.
The amount of memory in PC installed
Endoimage2 decreased to the alert
level of insufficient space.
Failed to transfer data to Endoimage2.
Setting value of IP address conflicts.
Failed to save a movie.
There is no data file to load in an image
record media in spite of [Configuration
load].
When OE was changed into Mode1 or
Mode2, the mode change failed.
Because of the abnormalities occurred
in communication with an endoscope
when changed OE into Off, the mode
change of OE failed.
It has reached the continuous-capture
ceiling of the memory capacity.
Solution
Restart the processor.
When the same message is displayed
repeatedly, contact a PENTAX Medical
service facility.
Contact a PENTAX Medical service
facility.
Contact a PENTAX Medical service
facility.
Try copying the data again.
Retry to transfer data after confirming
network connection.
Confirm the connection with a PC
activating Endoimage2.
Connect the processor to a PC with
sufficient memory.
Connect the processor to a PC with
sufficient memory.
Contact a PENTAX Medical service
facility.
Set the IP address not to conflict and
reboot the processor.
Confirm the correct connection with an
image record media or reboot the
processor.
If problem is not resolved after the above
procedure, contact a PENTAX Medical
service facility.
When image processing level is set up
highly, an error may occur to the
compression codec technology.
In that case, use setting it up with the
compression ratio for High quality or
Standard quality.
Confirm the existing file to load in an
image record media.
Perform the mode change again.
Reconnect the endoscope.
Try capture again after a while.