Ambient Temperature; Capacitor Life Expectancy - MAGSTIM RAPID2 Operating Manual

Hide thumbs Also See for RAPID2:
Table of Contents

Advertisement

Magstim Rapid
2
MEPPod
Class Type:
Channels:
CMMR:
Dynainic Range:
Amplitude accuracy:
Display resolution:
Cursor resolution:
Ti111ebase accuracy:
Timebase resolution:
CALIBRATION
Calibration is not required.
Class I, BF
2
>9OdB
± 30mV ±10%
±
10%
at
1
/2
full
scale
<2~0
<2~:()
<± 1 % of sample period
0.33% of sample period
10.3 Al"1BIENT TEMPERATURE
Permissible EnYinmment Conditions for Transport
Ambient Temperature Range:
Relative Humidity Range:
Atmosphere Pressure Range:
-l 9°C to 6O°C
10%,
to 80% Non Condensing
5 OkP a to !06kPa
Storage and Operating Conditions:
Operating T
~111perature
Range:
Storage Te111perature Range:
Coil Temperature Range:
5°C
to 4O°C
-19°Cto 60°C
5°C
to 4O°C
10.4 CAPACITOR LIFE EXPECTANCY
Life expectancy:
2 x 10
6
discharges at 70%, power level
8 x 10
5
discharges at 80%, power level
4 x 10
5
discharges at 90%, power level
2 x l 0
5
discharges at 100%, power level
10.5 OUTPUT SAFETY, REPETITION
Ri\
TE OF STilVIULUS:
3576-23-09
The following Table is taken from Eric
}~f.
Wassermann's paper - Risk and Safety of Repetitive
Transcranial Magnetic Stimulation: Report and recommendations from the International workshop on
the safety of Repetitive Tnmscranial Magnetic Stimulation June 5-7, l 996. It gives the maximum safe
duration (in second~) o_f single trains of
rTi.lfS
based on the
NINDS
experience.
Frequency
Intensity(% ofl\IEP threshold)
Hz
100
110
120
130
140
150
160
170
180
190
200
210
1
>1800
>180
360
>50
>50
>50
>50
27
11
11
8
7
0
5
>10
>10
>10
>10
7.6
5.2
3_6
2.6
2.4
1.6
1.4
1.6
10
>5
>5
4.2
2.9
1.3
0.8
0_9
0.8
0.5
0.6
0.4
0.3
20
2_05
1.6
1.0
0_55
0.35
0.25
0.25
0.15
0.2
0.25
0_2
0.1
25
1.28
0.84
0.4
0.24
0.2
0.24
0_2
0.12
0.08
0_12
0.12
0.08
Numbers preceded by the symbol> are the longest durations tested. No after discharge or spread of
excitation has been encountered v,rith single trains of rTMS at these combinations of stimulus frequency
and intensity.
Magnetic Field:
Pulse Characteristics:
The magnetic field produced by the Magstim Rapid
2
depends on the type
of coil connected. In the case of the Double 70mm Coil supplied with the
system, the field output is l.2T at the smface of the coil.
Biphasic Single Cosine Cycle - 400µs period
The MAGSTIM Company LTD
48
November 2009
220
6
1.2
0.3
0.1
0.08