Set-Up - MAGSTIM RAPID2 Operating Manual

Hide thumbs Also See for RAPID2:
Table of Contents

Advertisement

Magstim Rapid
2
3.6 SET-UP
1.
2.
3.
4.
5.
6.
7.
8.
9.
3576-23-09
Before using the instnunent ensure that the co1Tect fuses are fitted and that the
voltage selector is set to suit the country's supply voltage (this statement applies to
the Rapict2 Main Frame only). There are
2
fuses, rated at T 1.25A for both l 15V-
±10%, and 230V - ±10';,·o. For infonnation on changing the fuses, please see Section
9.
Place the system in a room near multiple power sockets and ensure easy access to
and around the equipment.
Label the entrances to the room to warn of the presence of strong magnetic fields and
to exclude those wearing pacemakers and/or electronic implants.
Connect the power cables provided to the power supply socket(s) on the rear of the
PSU. The power reqnirement of the Magstim Rapid
2
depends on the PSU used and
the parameters set by the UI.
The power cables must be connected dkectly to wall power outlet sockets, and
not via a terminal block. The Rapid
2
System must be used only with the supplied
mains leads fitted with an integral filter, as they are required to maintain the
System's compliance with EN 60601-1-2 regarding Electro-Magnetic emissions, as is
the earth strap fitted benveen the stimulator and the power supply unit (PSU).
Connect the HV cable from the PSU to the rear panel of the Main Frame. Connect
the mains link between the PSU Module and the Main Frame.
Connect the UI Controller to the Magstim Rapid
2
via the interface cable supplied.
Connect the Stimulating Coil to the front panel connector of the Magstim Rapid
2
.
Apply mains power to the Rapid
2
System by switching the mains power switch on
the rear panel of the Rapid
2
Main Frame to the ON position.
The MAGSTIM Company LTD
10
November 2009