User Maintenance; Technical Maintenance - MAGSTIM RAPID2 Operating Manual

Hide thumbs Also See for RAPID2:
Table of Contents

Advertisement

Magstim Rapid
2
3576-23-09
9.2 USER MAINTENANCE
At the start of each session the Operator must check the Magstim Rapid2,
PSU, UI ,md the stimulating coil for any signs of damage, paying particular
attention to the plastic casing, Ifthere are any signs of physical damage to
the Magnetic Stimulator, PSU, UI or the stimulating coil, they must not be
used,
The cooling vents should also be checked to ensure that they have not
become blocked, or obstrncted,
The UI contains 2 AA cell batteries, These are not rechargeable, and will
need to be replaced if patient inf<mnation is stored on the system, It is
important that regular checks are cmTied out to ensure that the batteries are
replaced ·when necessary. Failure to do so will result in the infonnation
being lost. The Battery voltage can be viewed in the User Configuration
Screen.
9.3 TECHNICAL MAINTENANCE
BACKUP BATTERIES
The coil pins should be checked before each session for any signs of pitting,
or burning, as under conditions of exceptionally hard use at high energy
levels, it is possible for the localised heating to manifest itself in the form
of micro-welds. Continued used in this condition will eventually result in
the coil pins/sockets becoming totally eroded and open circuit
Note: The pin burning is communicable - any coil with good pins that are
used on a stimulator with burned socket pins ·will have its pins damaged
immediately. The reverse is also true; a socket with good pins that has a
coil with burnt pins connected to it will have its pins damaged
immediately. If damage is noticed on any coil connector or stimulator coil
socket the complete system must not be used until all pins and sockets are
carefully examined for any damage. If any contacts show damage, even if
slight, they will need to be changed. If this is not done thoroughly there is
the risk that the cycle of contact damage will continue.
As contact repair is a specialised procedure, it is recommended that contact
replacement is undertaken by the Magstim Company, or one of its
authorised service centres. If additional advice or information is required,
please contact The Magstim Company.
Battery status can be checked on the user configuration screen and when
low, is reported on the main Options Menu screen.
The MAGSTIM Company LTD
44
November 2009

Advertisement

Table of Contents
loading

Table of Contents