MAGSTIM RAPID2 Operating Manual page 59

Hide thumbs Also See for RAPID2:
Table of Contents

Advertisement

Magstim Rapid
2
3576-23-09
Guidance and Manufacturer's Declaration - Electroma~netic Immunity
The Magstim Rapict1 System is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is used in such an environment.
Immunity Test
IEC 60601
Compliance
Electromagnetic enYironment -
Test LeYel
LeYel
~uidance
Electrostatic
± 6 kV contact
Meets requirement
Floors should be wood. concrete or ceramic
discharge (ESD)
±
8 kV air
tile. If floors are covered with synthetic
material, the relative. humidity should be at
IEC 61000-4-4
least 30%.
Electrical fast
± 2 kV for power
Meets requirement
Mains power quality should be that of a
Transient/ burst
supply lines
(Resetting of the
typical commercial or hospital
MEP Pod may be
environme.nt.
IEC 61000-4-11
± l kV fr,r input/
experienced during
output lines
single shot mode
under these
conditions.)
Surge
±
l kV differential
Differential mode
Anti surge protection needs to be
mode
compliant.
incorporated into the main supply to the
IEC 61000-4-5
equipment if surge protection is to be
± 2 kV common
Common mode
guaranteed.
mode
compliant to± lkV
only.
Voltage dips,
<5
~:Q
l/T
Meets requirement
Maim power quality should be that of a
short inte1n1ptions
(>95 ?o dip in UT) for
typical commercial or hospital
and voltage.
0.5 cycle
environment.
If
the user of the Magstim
Yariations on
Rapid
2
System requires continued
power supply
40 ?-o Uy
operation during power interruptions, it is
input lines.
(60
%
dip in UT) for
recommended that the Magstim Rapid
2
5 cycles
System be pmvered from an interruptible
IEC 61000-4-11
power supply.
70% Uy
(30 % dip in UT) for
25 cycles
<5% UT
(>95
%
dip in UT) for
5
sec
Power Frequency
3 A/m 50Hz
Meets requirement
Power frequency magnetic fields should be
(50!60Hz)
at levels characteristic of a typical location
Magnetic Field
in a typical commercial or hospital
environme.nt
IEC 61000-4-8
Radiated RF
80MHz - 2.5GHz
Coil adapter P/N
Equipment should only be used in the
Immunity
2Hz 80'% amplimde
3110-00 and Cooled
vicinity of other equipment compliant with
EN6 l 000-4-3
modulation
Coil
PIN
3380-00
EN60601-1-2.
between 80l'vIHz and
l00MHzmay
produce system
effors setting system
into safo mode.
Conducted RF
0.15MHz - 80MHz
Coil adapter P/N
Equipment should only be used in the
Immunity
2Hz 80%, amplitude
3110-00 and Cooled
vicinity of other equipment compliant with
EN61000-4-6
modulation
Coil
PIN
3380-00
EN60601-1-2.
between 30MHz and
80MHz may produce
system errors se-tting
system into safe.
mode.
NOTE
UT is the a.c. mains voltage prior to application of the test leve.l
The l\'IAGSTIM
Company
LTD
56
November 2009