3576-23-09
Note 1. Line only added to files by UI software version V6.0 or greater.
File line Details
The following details the content and data range limits for each data lines contained in the Magstim UI
generated file.
Version. This number is created by the UI and must be present in the file and ''must not be altered".
An invalid number ,vill result in the file not being imported correctly.
File. This is the file name used on the UI and must contain at least one character and not more than 18
characters. The file name can contain any standard alphanumeric character except a comma.
Status. This number is used by the UI to detennine what state the file was in when it was exported and
"must not be altered". An invalid number will result in the file not being imported correctly.
File Time. This line defines when the file was created on the
UI.
The range for each value is a follows:
seconds0-59,minutes0-59,hours 1-23,day 1-31,month 1-12,year 1970-2100.
Name. This is the name that appears in the patient data name windmv on the UI and must not exceed
40 characters. The name can contain any character except a comma.
Sex. This is the character that appears in the patient data sex window and must not exceed 1 character.
The character can be any character except a comma.
ID#. This is the string that appears in the patient ID window on the UI and must not exceed 18
characters. The string can contain any character except a comma.
Date of birth. This is the string that appears in the patient date of birth window on the UI and must not
exceed 18 characters. The date string can contain any character except a comma.
Protocol desniption. This is the string that appears in the patient protocol description window on the
UI and must not exceed 46 characters. The string can contain any character except a comma.
Comments. This is the string that appears in the patient comments window on the UI and must not
exceed 2 73 characters. The string can contain any character except a comma.
Trnin. This line contains the parameters displayed on the repetitive set up screen on the UI, (Standard
mode). The individual parameter value ranges are as follows:
Power
O -
100% at a resolution of 1
%,.
Frequency I - 100 Hz at a resolution of lHz.
Note the settable power to frequency values is deternlined by the system configuration, see user manual
for Rapid system for more details.
Duration 0.1 -
10
.0 seconds at a resolution of 0.1 seconds.
Number of pulses 1 - I 000 at a resolution of 1.
Wait time 1.0 - 60.0 seconds at a resolution of 0.1 seconds.
Number of trains 1 - 250 at a resolntion of
1.
Note all time periods are specified in multiples of tenths of a second i.e. if a time period of 6.5 seconds
is required then a value of 65 is entered between the commas.
Burst. This line contains the parameters displayed on the repetitive set up screen on the UL (Burst
mode). The individual parameter value ranges are as follows:
Power
O -
l 00~--,, at a resolution of l
%.
Frequency I - 100 Hz at a resolution of lHz.
Note the settable power to frequency values is determined by the system configuration, see user manual
for Rapid system for more details.
Number of pulses I - 10 at a resolution of I.
Bnrst frequency l - 10 Hz at a resolution of I Hz.
Number of bursts 1 - 500 at a resolution of 1.
Cycle time 1.0 - 500.0 seconds at a resolution of 0.1 seconds.
Number of cycles 1 - 250 at a resolution of I.
Note all time periods are specified in multiples of tenths of a second i.e. if a time period of 6.5 seconds
is required then a value of 65 is entered betv,reen the commas.
Session x. These lines contain the parameters displayed on the session set up screen on the UL The
individual parameter value ranges for line 1 are the same for all 30 lines and are as follows: