MAGSTIM RAPID2 Operating Manual page 14

Hide thumbs Also See for RAPID2:
Table of Contents

Advertisement

Magstim Rapid
2
3576-23-09
3.7 USER INTERFACE (UI)
ARM Button
0
TRIGGER Button
In\
v;w
ARM Button
.~++±+++-i:--.;.;.;...;...
SYSTEM STOP
Button
Output Control Knob
The Magstim Rapid
2
can be put into the ARMED mode by momentarily pushing the
Green RUN button. This can only be achieved if the stimulating coil is connected to the
COIL OUTPUT socket.
The Magstim Rapid
2
can be triggered, and a magnetic pulse produced, by pushing the
Yellow TRIGGER button.
SYSTEM STOP Button
0
The Magstim Rapid
2
can be put into the safe inactive default mode by momentarily
pushing the Red STOP button. In this mode, the instrument will discharge internally.
OUTPUT CONTROL Knob
This control allows the user to change the screen-selected parameter to the desired value.
It is therefore used to alter screen settings as well as cha11ging the power output of the
stimulator.
TOUCH-SENSITIVE SCREEN/ SYSTEM DISPLAY
All selections are made via the touch-screen. To select, touch the centre of the button
symbol next to the desired menu option. Do not press hard, or use a sharp or pointed
object to make the selection, as this may damage the touch screen.
In the set-up screens, all selectable items are coloured in pale blue; white items are non-
selectable. The Rotary Control Knob can only be used to change parameters already
selected. Details regarding the contents a11d operation of the UI Screens are given in
Sections 4, 5 and 6
Note: If the UI is not used for 30 minutes it will go into a standby condition.
In
standby, the screen ,vill appem blank a11d the blue LED in the upper right corner of the
front bezel will pulse. Touch the screen to restore the UI.
The MAGSTIM Company LTD
11
November 2009