4.5 REPETITIVE MODE: STANDARD SETUP
The Repetitive Setup Screen allows the user to set the following adjustable parameters to the desired
values to enable a session to be nm. If Repetitive Setup is accessed from Single Pulse Mode, the
protocol intensity value established in Single Pulse l\fode will be carried over into the power window.
Power
Frequency
Duration
Number
of Pulses
Wait Time
No. of Trains
Information
\Vindow
Print
Save Session
Options Menu
Select the power window and use the Rotary Control Knob to adjust the System
Power Level. The power can be increased, or decreased, in 1 % increments.
Selecting the Frequency ·window allows the user to set the frequency of the pulses to
a value between 0 and Frv1axHz in lHz increment5. The maximum frequency can be
up to lO0Hz depending on the power level set and the system configuration. A table
detailing the available pmver/ frequency combinations is included in Section 10.
Selecting the Duration window allows the user to set the duration of the pulse train to
between 0.1 and 10 seconds, in 0.1 second increments. On entering the duration and
frequency, the number of pulses is automatically calculated by the UI and displayed.
Selecting the Number of Pulses ·window allows the user to set the number of pulses in
a train to between 1 and 1000 in steps of l. The number of pulses cannot be set to
deliver more than 10 seconds stimulation. If the frequency is altered, the system ~Nill
automatically adjust the duration to fit the current number of pulses.
Selecting the '\Vait Time' allows the user to set the wait time to between l and 60
seconds, in 0.1 second steps. However, if the minimum wait time required for the
selected protocol is greater than that selected, the UI will display a message 'Wait
period too short, auto-set to minimum'. Select 'YES' to set the wait delay to the
minimum allowed for the chosen protocol. Select 'NO' to reset the protocol
parameters manually.
Selection of this option allows the user to set the number of times that the specified
train runs to a value between 1 and 250 in steps of
1.
Appears at the base of the screen. The Infonnation Window provides information
such as 'Coil OK to nm,' and ale11s the user to actions required, protocols that are
unable to be completed, or equipment which needs to be attached.
For Print operations see Section 6.
Save Session takes the user to the File Maintenance screen (See section 6)
Selection of this option will return the user to the Options Menu screen.
RUN SESSION This screen takes the user to the Repetitive Mode screen in order to run a session.