10.1 POWER
10.2 AMBIENT TEMPERATURE
10.3 CAPACITOR LIFE EXPECTANCY
Life expectancy:
10.4 OUTPUT
Repetition rate of stimulus:
The Magstim Company Limited 3001-23-04
Voltage:
Supply Frequency:
Power Input Fuse Rating:
Equipment should be used with the line cord provided. Where a fused
line cord is used, it is recommended that a 7A or 15A fuse be fitted for
220 – 240V and 110 – 120V operation respectively.
Max Power Requirements:
Standby
During discharge
Permissible Environment Conditions for Transport
Ambient Temperature Range:
Relative Humidity Range:
Atmosphere Pressure Range:
Storage and Operating Conditions:
Operating Temperature Range:
Storage Temperature Range:
Coil Temperature Range:
6
2 x 10
discharges at 70% power level
5
8 x 10
discharges at 80% power level
5
4 x 10
discharges at 90% power level
5
2 x 10
discharges at 100% power level
Once every 2 seconds at up to 0 - 49% power level
Once every 3 seconds at up to 50 - 79% power level
Once every 4 seconds at up to 80 - 100% power level
220-240V/110V-120Vac
50-60Hz
5A T x 2 (Europe; 220 – 240V)
10A T x 2 (USA; 110 – 120V)
100 VA
2000VA peak, 600VA average
-25ºC to 60ºC
10% to 80% Non Condensing
50kPa to 106kPa
5ºC to 40ºC
-25ºC to 60ºC
5ºC to 40ºC