4.4 SINGLE PULSE JVIODE (MEP POD CONNECTED.)
The MEP Pod must be connected to access the following options and to view the traces.
NOTE. The Rapid
2
must be switched off before connecting m· disconnecting the MEP Pod.
PARAMETERS PANEL
~:~=-~, 1 , _ 1, 1 , _ r, _ 1, _ 1_-_l_!_~_:. ! _[_:_r_:_:_:~;:~::::~~~~;:~;:::~i~~:~iilililli __ _
=~~,;a
fffflt::::::::::::::::::::::::::::::::ijfffi
·.·.·-~·-·································-·:·:•:•:•:•:•:•:•:•:•:•:•:•:•:•:•:·.·················
P1·ess the Parameters button (green when selected) to access the MEP Configuration panel.
Time Base
Selecting the Time Base window allows the user to set the MEP total sample
period. The selectable options are; 20m.s, 50m.s, l00ms, 200m.s and 500ms and are
selected using the Rotary Control Knob. The trace width will show the whole of
the selected sample period.
Filters
Selecting the Filters window allows the user to set the MEP Filter using the
Rotary Control
Knob. Selectable options: 2Hz-10kHz and 20Hz-10kHz.
Volts/Div Trace 1
Selecting the Volts/Div Trace window allows the user to set the scaling factor for
channel 1. Selectable options: S0µV, IO0µV, 200µV, 500µV, lmV, 2mV, SmV
and l0mV. The default setting is 200~lV.
Volts/ Di\' Trace 2 Selecting the Volts/Div Trace ,:vindow allows the user to set the scaling factor for
channel 2. Selectable options: 5011.V, IO0µV, 200µV, 500µV, lmV, 2mV, SmV
and l0mV. The default setting is 200~lV.