10.5 Calibration Procedures (CH1/CH2): DC mV/Volts
(measure)
Note: When recalibrating 'measure' functions for this
Perform the calibration as follows:
1. Connect the applicable calibration equipment specified
in Table 10-1; for example see
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Ref:
three-point calibration (-FS, Zero and +FS) for the
applicable set of ranges:
mV (measure) ranges
200 mV
2000 mV
4. Check the calibration is correct:
• Select the applicable Millivolts or Voltage (measure)
function (Ref:
• Apply the input values that are applicable to the
calibration:
mV: -1995, -1000, -195, -100, 0 (short circuit)
mV: 0, 100, 195, 1000, 1995
Volts (V): -30, -21, -20, -10, -5, 0 (short circuit)
Volts (V): 0, 5, 10, 20, 21, 30.
29 July 2016
range, any adjustment will affect the corresponding
source function calibration. The source function
must, therefore, be re-calibrated after adjusting the
measure function.
Section
116M5464 Revision -
Section
Section
Volts (measure) ranges
20 V
30 V
2.8).
Druck DPI620G-IS
2.8.
10.2) to do a
Page 221 of 248