• Check the error is within the limits specified in the
data sheet, Gauge & Absolute Range tables
values in the column 'Total uncertainty ... ').
The specified values include an allowance for
temperature changes, reading stability for one year,
and the uncertainty of the standard used for
calibration.
Ranges: gauge
≤ 700 mbar (10.0 psi)
> 700 mbar (10.0 psi)
†For a three-point calibration, do not apply more than -
Ranges: absolute
350 mbar (5.00 psi)
2 bar (30.0 psi)
7 bar (100.0 psi)
20 bar (300.0 psi)
≥ 350 bar (5000 psi)
29 July 2016
Table 10-15 Calibration pressures
90% of the specified FS for the unit.
116M5464 Revision -
Nominal applied pressure
mbar (psi)
-FS †
-FS
-900 (-13.1)
Nominal applied pressure
mbar (psi)
Zero
< 1.0 (0.02)
< 5.0 (0.07)
< 20.0 (0.29)
< 50.0 (0.73)
Use atmospheric
pressure as zero.
Druck DPI620G-IS
(use the
Zero
+FS
0
+FS
0
+FS
+FS
+FS
+FS
+FS
+FS
+FS