Introduction; Calibration Requirements - GE Sensing Druck ADTS 2 Series Calibration Manual

Air data test set
Hide thumbs Also See for Sensing Druck ADTS 2 Series:
Table of Contents

Advertisement

1
1

INTRODUCTION

The ADTS 2XX incorporates a calibration facility that operates through an enable switch,
located on the front panel under a calibration protection label. For the system to stay
accurate, a calibration check should be carried out at chosen intervals. If the accuracy
of the system is not within the specification, carry out a calibration adjustment.
Calibration process
The maintain calibration menu contains the PIN protected calibration adjustment
functions:
This manual describes the sensor calibration. An additional PIN protects the valve
calibration, as this procedure must only be carried out at the repair depot.
Test instrument characteristics
Pneumatic pressure
Sensor calibration
Valve calibration (not accessible)
T
1-1 C
ABLE
(1)
Range:
Accuracy:
Druck ADTS 2XX Calibration Manual
R
ALIBRATION
EQUIREMENTS
Performance specifications
(2)
35 mbar to
3000 mbar absolute
± 0.01% of reading
Test method
(3)
Compared against
pressure standard.
K0425 Issue No. 2