Example Procedure: Measure Or Source Current With Internal 15V Loop Power, Ch2 - GE DPI620G-IS User Manual

Intrinsically safe calibrator and communicator series
Table of Contents

Advertisement

User Manual
2.7
Example Procedure: Measure or Source Current with
Internal 15V Loop Power, CH2
Note: CH2 can be configured to measure or source
current using internal or external loop power. For an
example of external loop power, refer to
Safety Note: Internal 15V loop power is activated for CH2
functions measure and source current with internal 15V
loop power. With these set-ups an unregulated 15V
supply is presented across terminals '15Vo' and 'mA-
/COM' as shown in Figure
Figure 2-15 Connection Details for 15Vo Unregulated Current
Loop Output
29 July 2016
2-15.
116M5464 Revision -
Druck DPI620G-IS
Section
2.6.2.
Page 56 of 248