6.13.2

Method Example - Analog Trim

The Druck DPI620G-IS can perform an analogue trim on
the 4 to 20 mA loop without connecting to any external
reference meters.
1. Navigate to the calibration folder
2. Select Trim analog method.
3. Follow the on-screen instructions. If channel 2 is set to
Measure Current (15V), this can be used to supply the
reference meter value.
29 July 2016
Figure 6-11 Calibration Folder
116M5464 Revision -