GE DPI620G-IS User Manual page 224

Intrinsically safe calibrator and communicator series
Table of Contents

Advertisement

User Manual
Table 10-6 millivolts (source) error limits
Table 10-7 Voltage (source) error limits
29 July 2016
Source
uncertainty
(mV)
0
100
195
1000
1995
Source
uncertainty
(V)
0
3
6
9
12
116M5464 Revision -
Calibrator
(mV)
0.0001
0.00046
0.0009
0.003
0.006
Calibrator
(V)
0.000004
0.000019
0.000034
0.000049
0.000064
Druck DPI620G-IS
Permitted
DPI620G-IS
error (mV)
±0.0060
±0.0090
±0.0119
±0.0750
±0.0899
Permitted
DPI620G-IS
error (V)
±0.00048
±0.00060
±0.00072
±0.00084
±0.00096
Page 224 of 248

Advertisement

Table of Contents
loading

This manual is also suitable for:

Druck dpi620 genii

Table of Contents