 Check that the error is within the specified limits
(Ref: Table 10-4 or Table 10-5).
Table 10-4 Millivolts (measure) error limits
(short circuit)
Table 10-5 Voltage (measure) error limits
(short circuit)
29 July 2016
Applied
(mV)
±1995
±1000
±195
±100
0
Applied
(V)
±30
±21
±20
±10
±5
0
116M5464 Revision -
Calibrator
uncertainty
(mV)
0.051
0.040
0.0051
0.0040
0.0036
Calibrator
uncertainty
(V)
0.00052
0.00040
0.00031
0.00016
0.00008
0.000024
Druck DPI620G-IS
Permitted
DPI620G-IS
error (mV)
±0.0899
±0.0750
±0.0109
±0.0095
±0.0080
Permitted
DPI620G-IS
error (V)
±0.00195
±0.00172
±0.00090
±0.00065
±0.00052
±0.00040