10.3 Calibration Procedures (CH1/CH2): Current (measure)

Note: When recalibrating 'measure' functions for this
range, any adjustment will affect the corresponding
source function calibration. The source function
must, therefore, be re-calibrated after adjusting the
measure function.
Perform the calibration as follows:
1. Connect the applicable calibration equipment specified
in
Table
• CH1 (55 mA range) see
• CH2 (55 mA range) see
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Ref:
three-point calibration (-FS, Zero and +FS) for each
range: 20 mA and 55 mA.
4. Check the calibration is correct:
• Select the applicable Current (measure) function,
(Ref:
• Apply the following values:
• Check the error is within the specified limits (Ref:
Table 10-2).
29 July 2016
10-1; for example:
Section 2.6,
mA: -55, -25, -20, -10, -5, 0 (open circuit)
mA: 0, 5, 10, 20, 25, 55.
116M5464 Revision -
Section 2.6
Section 2.6
Section
2.7).