Sequence To Upload And Download File - GE DPI620G-IS User Manual

Intrinsically safe calibrator and communicator series
Table of Contents

Advertisement

User Manual
5.4.1

Sequence to Upload and Download File

Step
1. Connect Standard USB lead (as supplied) to the Druck.
DPI620G-IS calibrator.
2. Connect the lead to the USB port on the computer that
has the calibrator manager installed.
3. Use 4 Sight / Intecal v10 to set-up the procedure and
create a work order for the device.
The procedure includes the parameters for the
calibration, the number of test points, the relationship
and pass/fail tolerance.
4. Use the Download button in calibrator manager to
Download the file to the Druck DPI620G-IS calibrator. A
communications symbol will be displayed at the
bottom of the screen.
5. Select DASHBOARD >> DOCUMENTING >> RUN
PROCEDURE
6. In the Results window, select the filename specified in
calibrator manager.
7. Enter the User ID and the DUT Serial Number.
8. Tap on the Start button
necessary Channel options e.g. mA and Volts.
9. Use the Take Reading button at each point specified by
the procedure. A prompt shows for each point.
10. When all the readings are complete, tap on the Exit
Button
found/As Left).
11. To complete the process, use the calibrator manager
to Upload the file back into the 4 Sight / Intecal v10
database
29 July 2016
Procedure
. Look at the results on the display (As
116M5464 Revision -
. The Procedure sets up the
Druck DPI620G-IS
Page 104 of 248