Switch Test - GE DPI620G-IS User Manual

Intrinsically safe calibrator and communicator series
Table of Contents

Advertisement

User Manual
2.3.2

Switch Test

This utility is available with measure or source
functions.
The additional values displayed show signal values
(measure or source) when the instrument detects a switch
opening and closing. The difference between the two
values is displayed as hysteresis value for the switch. This
utility can be used with Ramp Automation, where the rising
signal causes the switch to change state and the falling
signal causes the switch to resume its' original state.
29 July 2016
Figure 2-8 Switch Test Example