Page 1
Sensing Druck DPI 880 Multi-function calibrator User manual - K405 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com...
Introduction Safety The DPI 880 Multi-function Calibrator is part of the Before you use the instrument, make sure that you read Druck DPI 8xx series of hand held instruments. and understand all the related data. This includes: all local safety procedures, the instructions for the UMM (if This series of instruments uses Intelligent Digital Output applicable), and this publication.
To start To start - Key to figure A3 (Display) Item Description To start - Key to figure A1/A2 (Instrument) Task indication for the switch test. Item Description = switch closed = switch open ❍ On or off button. UPM only. Task indication for the leak test. Left-hand soft-key.
To start - Prepare the instrument To start - Select a task (Measure and/or supply) Before you use the instrument for the first time: When the instrument is set up (Table 1), use the task selection menu to select the applicable task. •...
Table 3: Menu options - Task selections Table 4: Menu options - Settings (Input) ➁ ) (Dual Function, area Options Description (If applicable) Options Description ... Units Pressure Units (UPM only). If you select an IDOS task (If applicable) (Table 2/3). Select one of the fixed units of White button = A Dual Function is set.
Table 5: (Part of table) Menu options - Settings (Output) Operation Options Description This section gives examples of how to connect and use the (If applicable) instrument. Before you start: To select and set up values for the “Span Check” •...
Operation - Measure/supply mA To measure/supply a current: 1. Connect the instrument (Figure 1, 2 or 3) and, if necessary, adjust the Set Up (Table 1). 2. Select the task from the task selection menu (Table 2/3). ➁ ) to do two operations Note: Use the Dual Function area ( ➁...
Operation - Measure/supply Volts or mV Operation - Measure/supply Hz or pulses To measure/supply Volts or mV: To measure/supply Hz or pulses: 1. Connect the instrument (Figure 4/5) and, if necessary, 1. Connect the instrument (Figure 6) and, if necessary, adjust the Set Up (Table 1).
Operation - Measure/simulate an RTD or Ohms To measure/simulate RTD values or Ohms: 1. Connect the instrument (Figure 7/8) and, if necessary, adjust the Set Up (Table 1). 2. Select the task from the task selection menu (Table 2): 3. If necessary, adjust the Settings (Table 4/5) and/or the output values to the system (Table 7).
Operation - Measure/simulate a Thermocouple Operation - Transmitter calibration To measure/simulate the TC values: To calibrate a transmitter: 1. Connect the instrument (Figure 9) and, if necessary, 1. Connect the instrument (Figure 10/11) and, if adjust the Set Up (Table 1). necessary, adjust the Set Up (Table 1).
Operation - Switch test Operation - UPM Pressure measurements To do tests on a switch: Read all the instructions supplied with the UPM and then use the specified procedures to connect it (Figure 13/14). 1. Connect the instrument (Figure 12) and, if necessary, adjust the Set Up (Table 1).
Maintenance This section gives procedures to maintain the unit in a good condition. Return the instrument to the manufacturer or an approved service agent for all repairs. Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment.
Calibration Before you start the calibration, make sure that the time and date on the instrument are correct (Table 1). Note: GE can provide a calibration service that is traceable Selection sequence: to international standards. ➤ ➤ ➤ ➤ Task selection menu...
Calibration - Procedures: mA input Table 11: mA output error limits Output Calibrator Permitted 1. Connect the instrument to the calibration equipment error DPI 880 error (Figure 3). (mA) (mA) 2. Let the equipment get to a stable temperature 0.000 006 0.001 (minimum: 5 minutes since the last power on).
Calibration - Procedures: mV/Volts output 3. Set up the equipment with these conditions: 1. Connect the instrument to the calibration equipment Frequency meter: Gate time = one second (Figure 4). Signal generator: Output = 10V, unipolar, 2. Let the equipment get to a stable temperature square wave (minimum: 5 minutes since the last power on).
Calibration - Procedures: CJ input Calibration - Procedures: RTD (Ohms) output 1. Connect the instrument to the calibration equipment 1. Connect the instrument to the calibration equipment (Figure 9). (Figure 8). 2. Let the equipment get to a stable temperature 2.
Calibration - Procedures: TC (mV) input/output Specification 1. Connect the instrument to the calibration equipment: All accuracy statements include one year stability. • TC (mV) input = Figure 9b Specification - General • TC (mV) output = Figure 9d Languages English [Default] 2.
Specification - Temperature ranges (RTD) RTD type Standard Range °F Range °C Accuracy °F * Accuracy °C * Pt50 (385) IEC 751 -328 to 1 562 -200 to 0.90 0.50 Pt100 (385) IEC 751 -328 to 1 562 -200 to 0.45 0.25 Pt200 (385)
Specification - Temperature ranges (TC) Thermocouple type Standard Range °F Range °C Accuracy °F * Accuracy °C * IEC 584 -454 -328 -270 to -200 IEC 584 -328 2 502 -200 to 1 372 IEC 584 -346 2 192 -210 to 1 200 IEC 584 -454...