Druck ADTS 505 User Manual
Creating Custom Test Sequences for the ADTS 505
Instructions for generating or editing source data for test table values
The data shown on each screen or table of the test sequence comes from a single data source file. This .CSV
format file uses numeric data fields separated by commas. Files of this type may be viewed and edited using
a spread sheet program such as Excel. Files would normally be saved with an applicable name for easy
identification and selection.
A template file (TEMPLATE.CSV), provided on the software distribution with this product, may be used as the
basis for creating new aircraft test sequences. An example, below, shows how the data fields of the file can
be used in the test screens.
Test Table Title
Header 2
Header 1
Ref
Test Descriptor
Message
F
6-11 C
IGURE
USTOMER
T
S
F
S
EST
EQUENCE
ILE
ELECTIONS
6 - 31
Rate Units
Aim Units
Ps, Pt Aims
Ps, Pt Rates