Sequence To Upload And Download File - GE Druck DPI 620 User Manual

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620:
Table of Contents

Advertisement

7.3.1 Sequence to upload and download file

[EN] English - K0449
Step
Procedure
1.
P/N IO620-USB-RS232. Connect the USB Type A connector
to the DPI 620 calibrator; see
2.
Connect the RS232 to the serial communications port on the
computer that has Intecal installed; see
3.
Use Intecal to set up the Procedure and create a Work Order
for the Device.
The procedure includes the parameters for the calibration, the
number of test points, the relationship (for example: linear)
pass/fail tolerance.
4.
Use the Download button in Intecal to download the file to the
DPI 620 calibrator. During this process, there is a
communications symbol at the bottom of the display.
5.
6.
In the Results window that opens, select the Filename you
specified in Intecal.
7.
Enter your User ID and the DUT Serial Number.
8.
Tap on the Start button. The Procedure sets up the necessary
Channel options (for example: mA and Volts).
9.
Use the Take Reading button at each point specified by the
procedure. You see a prompt for each point.
10.
When all the readings are complete, tap on the Exit button.
You can then look at the results on the display (As Found/As
Left).
11.
To complete the process, use Intecal to Upload the file back
into the Intecal database.
Chapter
5.
Chapter
Tap the Run Procedure option
Documenting functions 7-3
Issue 1
5.