Sequence To Upload And Download File - GE Druck DPI612 User Manual

Portable pressure calibrator
Hide thumbs Also See for Druck DPI612:
Table of Contents

Advertisement

User Manual
4.4.1

Sequence to Upload and Download file

Step
Page 85 of 105
A Druck DPI612 calibrator device driver available as a
download from the website www.gemeasurement.com.
1
Connect the USB cable (See Section 1.10, Accessories)
to the Druck DPI612 calibrator.
2
Connect the lead to the USB port on the computer
that has the calibrator manager installed.
3
Use 4 Sight to set-up the procedure and create a
work order for the device.
The procedure includes the parameters for the
calibration, the number of test points, the relationship
and pass/fail tolerance.
4
Use the Download button in calibrator manager to
Download the file to the Druck DPI612 calibrator. A
communications symbol will be displayed at the
bottom of the screen.
5
Select:
DASHBOARD >> DOCUMENTING >> RUN PROCEDURE
6
In the Results window, select the filename specified in
calibrator manager.
7
Enter the User ID and the DUT Serial Number.
8
Tap on the Start button
necessary Channel options e.g. mA and Volts.
9
Use the Take Reading button at each point specified
by the procedure. A prompt show for each point.
10
When all the readings are complete, tap on the Exit
Button
(As found/As Left).
11
To complete the process, use the calibrator manager
to Upload the file back into the 4 Sight database.
Table 4-1
Procedure
. Look at the results on the display
[EN] English – 109M4017 Revision -
Druck DPI612