30-ampere
Circuit
Protector
30-ampere
Circuit
Protector
Earth
Ground
To ensure safety, the instrument should be wired to a remote emergency
switch (preferably outside the room where the centrifuge is housed, or
adjacent to the exit from that room), in order to disconnect the instrument
from the main power source in case of a malfunction.
Wire Insulation Color
Green/Yellow
Light Blue
Brown
Wall Outlet:
Hubbell 9330,
Bryant 96-30-FR,
or Equivalent
(NEMA 6-30 R)
Required Wire Connections
Terminal
Harmonized
Earth ground
Neutral
N
Live or Line
L
Measured
Line Voltage
North American Plug
Symbol
North American
L
L