Operation

Manual Operation

Manual Operation
NOTE
When using multiwell-plate carriers, enter the appropriate rotor, either SX4750 or SX4750a. The
speed will be automatically limited to 4000 RPM. (The 4000 RPM limit is valid when running up to three
stacked Beckman Coulter plates separated by cap strips. Reduce speed to 2700 RPM if running four
stacked plates).
Set rotor speed using RPM values only. To set speed to a specific RCF, use the RPM-to-RCF conversion
chart (GX-TB-015) to determine the applicable RPM to enter.
1
When a run-parameter key (
parameter can be entered or changed.
• Entry is completed when
2
To change an entry before you've pressed
different value.
a.
Press
b. To change an entry after you've pressed
3
If an unacceptable value is entered, an out of range message is displayed in the message line.
a. Enter the correct value.

Selecting a Rotor

1
Press
ROTOR
2
Press the next run-parameter key or press

Entering Run Speed

Enter a run speed up to the maximum speed of the rotor in use. Or, enter a relative centrifugal field
(RCF) value up to the maximum achievable RCF of the rotor.
1
Press
RPM/RCF
modes.
• Allowed speed range for the selected rotor is displayed.
2-4
ROTOR
ENTER/SAVE
to abort changes and start over.
CE
, then use the arrow keys or press
once to enter speed in RPM. Press
,
,
,
,
RPM/RCF
TIME
TEMP
ACCEL
or another function key is pressed.
or another parameter key, enter a
ENTER/SAVE
, press the run-parameter key again.
ENTER/SAVE
again to select a rotor.
ROTOR
.
ENTER/SAVE
again to toggle between
RPM/RCF
, or
) is pressed, a
DECEL
and
RPM
PN B07493AD
RCF