Beckman Coulter Optima XL Series Instruction Manual page 82

Preparative ultracentrifuge
Table of Contents

Advertisement

ESP™ Efficient Sedimentation Program
7-8
CURRENT VALUES
SPEED
(RPM)
TIME
REMAINING
(HR:MIN)
TEMP
(DEG C)
VACUUM: OFF
Figure 7-8. Using the ESP Settings
Once selected, the ESP settings usually cannot be changed. However, after the
ESP run settings are entered and you return to the Main Screen, you can select
a DECEL profile for the last step of the run as follows.
✭✮✬✮✵❈✹✻✸✯✲✵✮✼❛
1. Press the
2. Enter a profile number and press
in the lower left corner of the SETTINGS window. The DECEL profile
will be used during the last step of the ESP run.
The run terminates automatically at the completion of the final step unless
✼✽✸✹❛
you press
beforehand. (See Section 3 for general information on
multi-step programs.)
The ESP settings cannot be loaded for use if a run is already in progress.
✼✽✸✹❛
If
is pressed during an ESP or sequential run, the run will be termi-
nated. To resume and finish the run, press the
before the rotor comes to rest. The run will then resume from the beginning of
the step that was in progress when the
less of the time elapsed in that step before the run was terminated.
NOTE
If the run is interrupted by a loss of power, the run will
automatically resume and finish (as if there was no
power interruption) when the power is restored.
SETTINGS–ESP PROGRAMMED RUN
ACCEL:
1
DECEL:
9
softkey.
✮✷✽✮✻❛
. The selected profile will appear
✮✷✽✮✻❛
✼✽✸✹❛
hardkey was pressed, regard-
SPEED
(RPM)
TIME
(HR:MIN)
TEMP
(DEG C)
ROTOR
7
LOGGING:
TYPE 100 Ti
✼✽✪✻✽❛
and
hardkeys