Selecting a Rotor

1
Press
ROTOR.
The
Rotor:
bucket,
2
Press the appropriate soft key to select the rotor type.
A list of rotors of the selected type appears.
Swinging bucket rotors are displayed below.
a. To select a different rotor type, press
3
Select the rotor name by pressing the appropriate soft key.
a. Press
The selected rotor name appears in the

Entering Run Speed

Enter a run speed up to the maximum speed of the rotor in use (10,000 RPM maximum). Or, enter a
relative centrifugal field (RCF) value up to the maximum achievable RCF of the rotor.
PN J3HC-IM-5AD
field flashes, and rotor type abbreviation soft keys (
JLA:
lightweight fixed angle) appear.
Rotor : - -
Speed : 0
RPM
=
0 x G
Time :
00:00
Temp : 25°C
Max Temp : 29°C
Accel : MAX
Decel : MAX
ENTER.
JA:
fixed angle,
5.0
3.4A
4.2
ROTOR
again and then press another soft key.
Rotor:
field.
Operation
Manual Operation
ENTER
must be pressed
JS:
swinging
2-5