6.3. FIELD UPGRADING TO THE DMF VERSION

For field upgrading to the DMF (Dento-maxillo facial) version, follow
the instructions below.
(1)
Switch off the system.
(2)
Remove the panel with the mirror (a) by unscrewing the
screws (b) and disconnect the connector X31.
a
B-26
(3)
(4)
PC1
X31
C1 24
D33
b
RR 2
D37
D4 6
C1 53
(03.0)E
ORTHORALIX 9200
Remove from the smart key the integrated circuit (could be
mounted in the socket D2 or D3).
D1
D1
D2
D2
D3
D3
Mount the integrated circuit on the PC1 control board into
the socket D33 or D37 indifferently.
D2 8
D2 9
D3 0
D3 2
RR 1
D3 4
D3 5
D3 6
C1 33
R1 27
R1 28
R1 29
R1 30
RR 4
RR 3
C1 44
D4 4
D4 3
C1 49
RR 7
RR 6
R1 38
C1 54
R1 40
RR 8
X26
X28
X27
X29
Service Manual
NEW
D31
PC1
R3 00
R1 22
R1 24
R1 23
D3 8
D3 9
D4 0
D4 1
D4 2
RR 5
C1 52
R1 33
D45
D4 5
R1 39
R1 41
H5
X33
X31
X32
X30