Section F
Service Manual

5.4. REPLACING OPTOSWITCHES

Two types of optoswitches could be mounted in the system:
Type 1
looking at the optoswitch from the front it is possible to see two
letters printed on it: an E and a S.
For the replacement, the wires must be connected as shown in the
figure below:
REAR VIEW
GREEN
WHITE
E
BROWN
YELLOW
S
Type 2
Looking at the optoswitch from the front it is possible to see that
one part is grey and the other one is black.
For the replacement, the wires must be connected as shown in the
figure below:
REAR VIEW
GREY
GREEN
WHITE
BROWN
YELLOW
BLACK
No other special precautions are required for replacement of the
optoswitches, as they are mounted at fixed positions.
After replacement of the X, Y, or R movements optoswitches
proceed with the relevant cam adjustment, verifying the centering
of the system using the image quality phantom (see Section C and
Section F of this Service Manual).
If a collimator optoswitch is defective, it is necessary to replace
the optoswitch board PA2: verify then that the three positions of
the collimator (panoramic, transcan and ceph) are correctly
selected by the optoswitches.
F-34
(03.0)E