5.10. REPLACING THE PA5 AEC CONTROL BOARD

After the replacement of the PA5 AEC control board, follow the
instructions herewith given:
CASE (A): PAN PA5
(1)
verify the jumper configuration: see the figure below.
(2)
perform the pan AEC calibration using the service program
AEC slope (see the related paragraph in this Section F for
more details)
(3)
enter the pan AEC calibration parameter using the service
program AEC pan tune (see the related paragraph in this
Section F for more details)
(4)
record this calibration value in the Configuration Chart of
the Section C
(5)
record this calibration value also on the label glued close to
the PA5 board, erasing the old value (C).
CASE (B): CEPH PA5
(1) verify the jumper configuration: see the figure below.
(2)
perform the ceph AEC calibration using the service program
AEC slope (see the related paragraph in this Section F for
more details)
(3)
enter the ceph AEC calibration parameter using the service
program AEC ceph tune (see the related paragraph in this
Section F for more details)
(4)
record this calibration value in the Configuration Chart of
the Section C
(5)
record this calibration value also on the label glued on the
lead plate placed on the rear side of the ceph cassette
carriage, erasing the old value (C).