On the display the following message will appear:
where the numbers X and Y may have the following values, corresponding to the different status of the inputs of the PQ1 converter board:
X = 1
when the tubehead is in the panoramic position
X = 2
when the tubehead is in the ceph position
X = 3
when rotating the tubehead from panoramic to ceph position or viceversa
Y = 8
when the collimator is in the transcan position and the panoramic cassette is inserted in the carriage
Y = 9
when the collimator is in the transcan position and the panoramic cassette is not inserted in the carriage
Y = A
when the collimator is in the panoramic position and the panoramic cassette is inserted in the carriage
Y = B
when the collimator is in the panoramic position and the panoramic cassette is not inserted in the carriage
Y = 4
when the collimator is in the ceph position and the panoramic cassette is inserted in the carriage
Y = 5
when the collimator is in the ceph position and the panoramic cassette is not inserted in the carriage
Y = 0
when moving the collimator from one position to another (i.e. from panoramic to ceph) and the panoramic cassette is inserted in
the carriage
Y = 1
when moving the collimator from one position to another (i.e. from panoramic to ceph) and the panoramic cassette is not inserted
in the carriage
DENTSPLY Gendex
CONVERTER INPUTS READ
CONV INPUT READ
XY 0 0 21