Host Input Read - DENTSPLY ORTHORALIX 9200 Manual

Table of Contents

Advertisement

Section F
On the display the following message will appear:
where the numbers X, Y, W and Z may have the following values, corresponding to the different status of the PC1 system control board.
X
NU
S4-2
OPTO C
OPTO R B
1
0
0 = no reset
1 = obsc
1
1
1= reset
0 = not
obsc
LEGENDA:
NU = NOT USED
S4-2 = dip switch S4-2 on the PC1 board
OPTO C = optoswitch used to identify the half run position of the panoramic cassette carriage
OPTO R B = optoswitch used to reduce the speed of the R movement
S4-1 = dip switch S4-1 on the PC1 board
OPTO R A = optoswitch used to identify the reset position of the R movement
BRAKE OFF UP = pushbutton to release the brakes pushed to the upper position
OPTO Y = optoswitch used to identify the reset position of the Y movement
CEPH CASS = microswitch used to identify wether the ceph cassette is inserted or not
DEAD MAN = X-ray command S4
BRAKE OFF DOWN = pushbutton to release the brakes pushed to the lower position
OPTO X = optoswitch used to identify the reset position of the X movement
OBSC = optoswitch obscured
EXAMPLES:
DISPLAY INDICATION
8E
8F
(8 - 1110)
(8 - 1111)
8E
8F
(8 - 1110)
(8 - 1111)
8E
8F
(8 - 1110)
(8 - 1111)
8E
8F
(8 - 1110)
(8 - 1111)
8E
8E
(8 - 1110)
(8 - 1110)
If the values displayed are not congruent with the table above, there is a malfunctioning of the optoswitches, of the wiring or of the PC1
board.
F-68

HOST INPUT READ

HOST INPUT READ
8X 8Y 8W 8Z
Y
NU
S4-1
NU
OPTO R A
1
0
1
0 = no reset
1
1
1
1= reset
8D
8D
(8 - 1101)
(8 - 1101)
8C
8D
(8 - 1100)
(8 - 1101)
89
8D
(8 - 1001)
(8 - 1101)
8D
89
(8 - 1101)
(8 - 1001)
8D
8D
(8 - 1101)
(8 - 1101)
ORTHORALIX 9200
W
NU
BRAKE OFF
OPTO Y
UP
1
0 = off up
0 = reset 0 = ins
1
1 = on
1= no
reset
Rotating arm in reset position
Ceph cassette not inserted
Rotating arm in reset position
Ceph cassette inserted
Rotating arm in reset position
Brake release pushbutton in the upper position
Rotating arm in reset position
Brake release pushbutton in the lower position
X-Y-C movements in reset position
R movement not in reset position
Ceph cassette not inserted
(03.0)E
Service Manual
Z
CASS
DEAD BRAKE OFF OPTO X
CEPH
MAN
DOWN
0 = on 0 = off
0 = reset
down
1 = not
1 = off 1 = on
1= no reset
ins
CONDITION
DENTSPLY Gendex
DEAD
MAN
0 = on
1 = off