Section F

4.3.10. Converter input read

With this service program you can visualize the status of the
hardware inputs of the converter board PQ1, with four hexadecimal
numbers.
In this way you can check whether the PH3 boards are recognizing
the right inputs signals.
The correspondence between the hexadecimal numbers and the
status of the hardware inputs is shown in page FZ-2.
4.3.11. Pan and Ceph PH3 input read
This service program is applicable only for Orthoralix 9200 DDE
versions.
With this service program you can visualize the status of the
hardware inputs of the panoramic PH3 and ceph PH3 boards, with
four hexadecimal numbers.
In this way you can check whether the converter board is
recognizing the right inputs signals.
The correspondence between the hexadecimal numbers and the
status of the hardware inputs is shown in page FZ-3.
The following message will appear on the display:
PH3 PAN : CEPH READ
X XX: X XX
The first two numbers are related to the panoramic PH3 board and
the second two are related to the ceph PH3 board.

4.3.12. Exposure counter

By selecting this service program you can visualize the number of
exposures made.
The following message will appear on the display:
EXPOSURE COUNTER
P=XXXXX
where the number after the P indicates the panoramic exposure
made and the number after the C the teleradiographic ones.

4.3.13. Software version.

With this service program you can check the software version
mounted on converter board PQ1, on host computer board PC1 and
on keyboard control board PC7.
The following message will appear on the display:
SOFTWARE VERSION
H=XX
where the number after H indicates the software version of the
host computer board PC1, after CV the software version of the
converter board PQ1 and after CS the software version of the
keyboard control board PC7.
F-22
C=XXXXX
CV=XX CS=XX
ORTHORALIX 9200

4.3.14. PA7/PH3 software version

This service program is applicable only for Orthoralix 9200 DDE
versions.
With this service program you can check the software version
mounted on the Auxiliary board PA7, on Panoramic DMU Interface
board PH3 and on Ceph DMU Interface board PH3.
The following message will appear on the display:
A=XX HP=XX HC=XX
where:
A )
the number after A indicates the software version of the
Auxiliary board PA7
B)
the number after HP indicates the software version of the
Panoramic DMU Interface board PH3
C)
the number after HC indicates the software version of the
Ceph DMU Interface board PH3.

4.3.15. Ceph arm

This service program is applicable only for Orthoralix 9200 DDE
versions.
This service program indicates if the ceph arm has been properly
recognized by the host computer board PC1.
The following message will appear on the display:
where < NO > is the default condition.
When the ceph arm is installed, at the first system switching on,
the host computer board PC1 is detecting the presence of the ceph
arm and automatically changes the setting from NO to YES.
Please take note that, if the ceph arm is removed, in this service
program will remain the setting YES: this can generate error
messages when the system is switched on. So, in this case,
modify manually the setting from YES to NO by pressing the keys
projection+ or projection- on the keyboard.

4.3.16. X-ray enable

With this service program you can have a complete X-ray cycle but
without X-ray emission.
This could be useful for example during an exhibition, or in any
other case where you require a complete simulation but without X-
ray.
The following message will appear on the display:
where < YES > is the default condition.
(1)
modify the condition (YES / NO) by pressing the keys
projection+ or projection- on the keyboard.
(2)
the new condition will be stored when the system is
switched off.
(03.0)E
Service Manual
PA7/PH3s SW.VER.
CEPH ARM Y/N
< NO >
XRAYS ENABLE Y/N
< YES >