5.4.

INSTALLING THE CEPH ARM

5.4.1. Procedure for Orthoralix 9200

To install the ceph arm with Cephalostat follow the procedure below
(applicable only for the CEPH versions):
(1)
Remove the right-hand cover (a) of the Z-carriage by
unscrewing five screws.
(2)
Prepare the ceph arm close to the Z-carriage.
(3)
Remove the metallic plate (i) by removing the four fixation
screws.
(4)
Feed the free cable connector (b) from the arm through the
hole provided and connect to socket X29 of the host com
puter board PC1. The wire must pass in the rear side of the
metal plate where the PC1 is mounted.
(5)
If the AEC option is present connect the connector X34A
from the ceph arm to the X34B of the Z-Carriage.
DENTSPLY Gendex
(6)
Fix the ceph arm (c) to the Z-carriage by means of the four
hexagonal head screws (M8x25) and rings supplied.
(7)
Remove the panel (h) from the right-hand cover (a).
(8)
Replace the right-hand cover (a) of the Z-carriage.
(9)
Mount the ear plug support (d) into the cephalostat by
means of the two screws (e). The ear plug support has
two vertical mounting positions: please ask to the user
which position in more convenient.
(10)
Mount the nasion rest (f) into the pin and fix it with knob (g).
(03.0)E
Section B
X34A-X34B
B-11