6.

UPGRADES

6.1.

FIELD UPGRADING TO THE CEPH VERSION

For field upgrading to the CEPH version refer to the relevant Service
Manual supplied with the Ceph arm for detailed installation
instructions.
6.2.
FIELD UPGRADING TO THE TRANSCAN
VERSION
For the field upgrading to the Transcan version, follow the
instructions below:
(1)
Switch off the system.
(2)
Remove the panel with the mirror (a) by unscrewing the
screws (b) and disconnect the connector X31.
a
DENTSPLY Gendex
PC1
X31
b
(03.0)E
(3)
Remove from the smart key the integrated circuit (could be
mounted in the socket D2 or D3).
(4)
Mount the integrated circuit on the PC1 control board into
the socket D33 or D37 indifferently.
D2 8
D2 9
D3 2
RR1
C1 24
D3 4
D3 5
D33
RR2
C1 33
R1 27
R1 28
R1 29
R1 30
D37
RR4
RR3
C1 44
D4 4
D4 3
C1 49
RR7
RR6
D4 6
C1 54
RR8
X26
C1 53
X28
X27
X29
Section B
D1
D1
D2
D2
D3
D3
D3 0
NEW
D31
PC1
R3 00
D3 6
R1 22
R1 23
D3 8
D3 9
D4 0
D4 1
D4 2
RR5
C1 52
R1 33
D45
R1 38
R1 39
R1 40
R1 41
H5
X33
X31
X32
X30
B-25
R1 24
D45