Mounting The Front Panel With Mirror - DENTSPLY ORTHORALIX 9200 Manual

Table of Contents

Advertisement

ORTHORALIX 9200
Service Manual
5.5.
MOUNTING THE FRONT PANEL WITH
MIRROR
The front panel with mirror is supplied disassembled from the
Z-carriage.
To mount it proceed as follows:
(1)
Turn the rotating arm to the 90° position and move it in order
to have adequate access to the electronic rack.
(2)
Fit the top of the front panel with mirror into the slot (b) of
the Z-carriage.
(3)
Connect the two centering lights (laser projectors) to
connector X31 of the system control board PC1.
(4)
Check the proper insertion of cover (d) into the springs (e).
(5)
Lastly, close the front panel and fix it by means of the two
Allen screws M5x15 (c) provided.
DENTSPLY Gendex
(03.0)E