Section B
5.6.

MAINS VOLTAGE PROGRAMMING

The Ortoralix 9200 system can operate with a nominal mains voltage
in the 115 VAC to 250 VAC range.
The voltage range is selected automatically by the internal electronic
circuit.
For the detailed mains requirements please refer to Section A
ELECTRICAL DATA.
5.7.

CONNECTION TO THE MAINS

The system is selected for a single phase mains supply.
If the mains supply available is less than 150 VAC, replace the
fuses F1-F2 10A on the mains filter board, with the 15A fuses
supplied.
B
e
o f
e r
o c
n n
c e
n i t
t g
t o
e h
m
a
n i
a
e r
p
e r
e s
. t n
B-14
W
R A
I N
G N
, s
n e
u s
e r
h t
t a
o n
d
n a
e g
o r
s u
v
CASE (A): SINGLE PHASE MAINS SUPPLY
For the connection to the mains proceed as follows:
(1)
Remove the cover (a) from the column (4 screws).
(2)
Feed the mains supply cable (b) through the hole provided
into the column.
(3)
Connect the hot wire (L), the neutral wire (N), and the
ground wire (Y/G) to the mains connection points (see
figure below) of the mains filter board.
CASE (B): DOUBLE PHASE MAINS SUPPLY
If only a double phase mains supply is available, proceed as follows:
(1)
Remove the cover (a) from the column (4 screws).
(2)
Feed the mains supply cable (b) through the hole provided
t l o
g a
s e
into the column.
(3)
Locate jumper W1 on the mains filter board: remove it in
order to insert the second mains fuse F2.
(4)
Connect the hot wire (L), the neutral wire (N), and the
ground wire (Y/G) to the mains connection points (see
figure below) of the mains filter board.
(03.0)E
ORTHORALIX 9200
Service Manual
DENTSPLY Gendex