6.4. ADJUSTING THE LIGHT PROJECTORS

W
D
i r u
g n
t
e h
a
j d
s u
m t
n e
p t
o r
e c
d
e r i
t c
t y l
t o
e h
a l
e s
b r
a e
. m
To adjust the horizontal light projector, proceed as follows:
(1)
Make sure that the system is in the reset position.
(2)
Insert the light phantom (a) (code 4519 190 00922) into
the bite block holder (b).
(3)
Switch on the light and verify that the horizontal light is
sharply depicted over the whole horizontal black line of
the phantom.
F-58
A
R
N
N I
G
u d
, e r
v a
i o
e d
o t
x e
o p
e s
t
e h
(4)
If the light beam is not horizontal, remove the mirror panel
(c) by unscrewing two Allen screws (d).
(5)
Disconnect the connector X31 from the PC1 control board.
y e
s e
C
(6)
Access to the horizontal light projector (e) and unlock
the screw (f): then rotate the light projector to correct
the horizontal position of the light beam.
(7)
Reinstall the mirror panel (c) and verify if the horizontal
light is adjusted (if not, repeat the procedure from point
(4) to (6)).
(03.0)E
ORTHORALIX 9200
Service Manual
D
DENTSPLY Gendex
X31