8.
INTRODUCTION TO PROGRAMMING PRESET BUTTONS:
By default, the Preset Buttons are programmed
to recall the file series shown in Chart 1 on
page 5.
The Preset Buttons are also used to navigate
the various menus in the System Setup and
File Library.
The Preset Buttons can be customized to suit
individual preferences. There are three ways to
program a preset button.
The Preset Buttons can be programmed with:
1. An Entire File Series from the File Library
2. A Single File from the File Library
*
3. A User Defined Preset
When recalling a preset button, three different
displays will appear depending on how the pre-
set button was programmed:
1. If an Entire File Series was programmed
on the preset button (as is the case for
the factory default setup), the File Series
and File Name will be displayed, as
shown in the example below.
File Series
GT SERIES 20, 30, 40
ACC 90, 70, 50, 35/12
File Name #1
Pressing the same preset button again will
step to the next file in the file series, as
shown below.
File Series
GT SERIES 20, 30, 40
GT YEL 20/10 & 20/08
File Name #2
*
NOTE: When PRO-POST DRILLS is selected, there will not be a prompt for "LOAD ENTIRE SERIES"
because there is only one selection.
2. If a Single File was programmed on the
preset button, the words "Single File
Preset" and the File Name will be dis-
played, as shown in the example below.
Preset Type
SINGLE FILE PRESET
PROTAPER S1 & SX
File Name
3. If a User Defined Preset was programmed
on the preset button, the words "User
Defined Preset" and the Handpiece Ratio,
Speed in RPM, and Torque in
Gram•Centimeters, will be displayed, as
shown in the example below.
Preset Type
USER DEFINED PRESET
1/8 AG
300
Ratio
Speed
4. For all programming modes:
When the Endo DTC
the foot pedal pressed), the Operation
Screen will be displayed. The Operation
Screen will show the Handpiece Ratio,
Speed (in RPM), Torque (in
Gram•Centimeters), Reverse Beeper (if
enabled), and Auto Stop Reverse (if
enabled).
Ratio
1/8 AG
ASR BPR
{Auto Stop Reverse}
{Reverse Beeper}
240
Torque
®
is in operation (with
Speed
Torque
300
240
RPM g-cm
9