Each alert has its own settings. All of those settings are explained here:
st
Delay 1
Alert (High alert only): Turn on to delay your first alert until your sensor reading is at
or past the high alert setting for a while. You choose how long.
For example, if you set Delay 1
or above your high alert level for 20 minutes before you get the alert.
For More Than: Don't get an alert until a system issue lasts this long. You choose how long.
Fall Rate: For the Falling Fast alert, you choose the sensor reading change rate:
2–3 mg/dL per minute or
l
3 mg/dL or more per minute
l
For example, if you turn this on, when your glucose falls fast, you'll get an alert.
You can add a glucose level to this. If you do, you'll get an alert when your glucose is at or below
that level and falling fast.
62
st
Alert to 20 minutes for your High alert, your glucose must be at