Agilent Technologies 7250 Troubleshooting And Maintenance Manual page 83

Accurate-mass quadrupole time-of-flight gc/ms system
Hide thumbs Also See for 7250:
Table of Contents

Advertisement

4
General Maintenance
This procedure demonstrates how to remove the necessary covers to access
the left side lifting handles. The handles on the right and rear sides of the
instrument are not covered, and require no special procedures to access.
Procedure
1 To remove the front left grill, disengage the two captive screws on the left
Captive screws
Figure 13.
2 Swing the grill open, and remove it by disengaging the cover tabs on the right
3 To remove the left side cover, disengage the two captive screws at the front
Captive screws
Figure 14.
7250 Accurate-Mass Q-TOF Troubleshooting and Maintenance Manual
side edge of the grill. (See
Left front grill captive screws
side.
edge of the cover. (See
Left side cover captive screws
Figure
13.)
Figure
14.)
83